Posterior capsule tear caused by misalignment within a silicone irrigation/aspiration tip

Marcı́lio G Barros, MDac, Robert H Osher, MDCorresponding Author Informationab

Accepted 12 November 2003.

Abstract 

We describe a surgical complication with the Alcon silicone irrigation/aspiration (I/A) tip. In 2 consecutive cases, the posterior capsule was torn because of a misalignment between the inner metallic shaft and the hole in the silicone I/A tip. To achieve the protective benefits of this device, the surgeon must visually confirm proper alignment of the outer silicone and inner metallic portion of the tip.